BACKGROUND:
Patients with vertebral column deformations are exposed to high risks associated with ionising radiation exposure. Risks are further increased due to the serial X-ray images that are needed to measure and asses their spinal deformation using Cobb or superimposition methods. Therefore, optimising such X-ray practice, via reducing dose whilst maintaining image quality, is a necessity.
OBJECTIVES:
With a specific focus on lateral thoraco-lumbar images for Cobb and superimposition measurements, this paper outlines a systematic procedure to the optimisation of X-ray practice.
METHODS:
Optimisation was conducted based on suitable image quality from minimal dose. Image quality was appraised using a visual-analogue-rating-scale, and Monte-Carlo modelling was used for dose estimation. The optimised X-ray practice was identified by imaging healthy normal-weight male adult living human volunteers.
RESULTS:
The optimised practice consisted of: anode towards the head, broad focus, no OID or grid, 80 kVp, 32 mAs and 130 cm SID.
CONCLUSION:
Images of suitable quality for laterally assessing spinal conditions using Cobb or superimposition measurements were produced from an effective dose of 0.05 mSv, which is 83% less than the average effective dose used in the UK for lateral thoracic/lumbar exposures. This optimisation procedure can be adopted and use for optimisation of other radiographic techniques.